Release Notes RAPTOR SERVER v1.17
Tenant configuration is now extended to allow additional MOD settings. Three new settings have been added:
Click Menu (on the top right) ➡️ Tenant-Admin ➡️ Configure Tenant

1) Force MOD: If this setting is true, a MOD must be selected when saving or approving a measurement.

2) Hide Global MOD: When this setting is true, global MODs are not displayed in the MOD drop-down list when editing a measurement.


These new settings also affect some API endpoints (see the API section below).
We have made some changes to make it easier for the user to search and export multiple measurements.
Click Menu (on the top right) ➡️ Exports ➡️ Export multiple measurements A single text field for searching multiple measurements (QR-Code/Product/Batch): Now the user can filter measurements using
the Product Code (two characters, e.g.: 02), the Batch (five characters, e.g.: 02ABC) or the QR-Code (eight characters, e.g.: 02ABC123)


These filtering parameters were also changed in the API. Please see the corresponding section below.
Excel exports of New and Approved measurements of ALEX², FOX, PAX now include two new columns to provide the same information as in the PDF report:

Also, the units have been added to the header of the values.
We have detected that in ALEX² and FOX the Number of tested allergen sources was wrong since our last update showing the Number of tested allergens instead.
We have fixed the issue and now the correct number will be shown in every report of a measurement without a MOD applied.

We have updated the Agents to the latest version of .NET.
In addition to including the latest security updates, these new Agent versions no longer require the .NET framework to be installed, resulting in a faster and easier setup routine.

Data collected in the field has shown that one allergen included in the QualityXplorer does not react according to previous data collected from in-house testing.
Bet v 6 showed abnormal behavior in several centers using QualityXplorer lot 2. Therefore, Bet v 6 has been removed from the list of allergens considered in the QualityXplorer.
The product composition remains unchanged, but the allergen was removed from the data points used to determine the performance of the assay.
In the event of an incident where the Shutdown button is unavailable on the Dashboard page, users can rest assured that a Shutdown option remains accessible on the MAX Configuration page.
To further enhance our software’s responsiveness in critical situations, we have introduced an additional feature. Now, users can select the “Shutdown (skip clean cycles)” option directly from the drop-down menu. This function is specifically designed for emergencies requiring the pipettor to maintain its current position without initiating clean cycles.
We recommend users familiarise themselves with this update to ensure swift action when necessary.

In our continuous efforts to optimise the workflow and accelerate operational procedures, we have made a significant update. The priming step after requesting to open the top cover has been eliminated.
This change is applicable when the MAX is in the standby state or the manage inventory state, streamlining the process and saving valuable time. However, it’s important to note that when requesting to open the top cover in the volume inventory state, the system will operate as before, initiating the priming step. This ensures that the precision and integrity of our processes are maintained where necessary.