Release Notes RAPTOR SERVER v1.21

Highlights

GUI – Improved QC Measurements Handling

  • Unified evaluation logic: The QC evaluation strategy has been harmonized. The same color highlighting rules now apply on both the New Measurements page and the Device QC Measurements list.
  • Barcode validation: Sample codes assigned after a run are no longer reclassified as QC measurements (e.g., 3100004, 3222670). Such samples will not appear in the QC list, will not be included in QC charts, and will not be highlighted in the New Measurements list.

  • ⚠️ Ensure that QC Sample barcodes are correctly scanned or recognized as QC -if manually filled out- by RAPTOR SERVER before the run starts.

  • Reprocessed QC measurements: When reprocessing a QC sample, the original capture timestamp is now consistently used to order results in both QC lists and QC charts. This guarantees stable sorting across all views.

GUI – CV Calculation Bug Fix

  • The formula used to calculate the Coefficient of Variation (CV) for each allergen in the QC Charts Trend View has been corrected.

  • CV is now consistently calculated as: CV = (Standard Deviation / Mean) × 100%

GUI – Export of Individual QC Charts

  • Each QC Chart in the Trend View can now be exported individually.
  • The QR code and measurement date are now stamped on every QC datapoint.
  • Supported export options:
    • Image file – download the chart as an image for documentation or reporting.
    • Datapoints – copy the numerical values for further processing or archiving.
v1-21_Image1


To access the QC section: Menu ➡️ Tenant Admin ➡️ Manage MAX devices/ImageXplorers ➡️ QC ➡️ Chart Trend View

Legacy/2025 Report - Report Template Duplication

New functionality: Existing report templates can now be duplicated. This allows admins to quickly create a new template based on an existing configuration, avoiding manual re-entry of settings.

v1-21_image2

Access the Report template configuration page: Menu ➡️ Tenant Admin ➡️ Manage Report Templates ➡️ Open existing template ➡️ Duplicate

2025 Report – Updates

  • Section title change: In the ALEX Air report, the group title “Insects & Venoms” has been renamed to “Insects” to more accurately reflect the content.

  • Address block & footer formatting: HTML is no longer supported in the address and footer fields. Line breaks can now only be created using indentation (Enter) within the text blocks.

v1-21_image3

  • Device identification: Reports can now display the device type and serial number for each measurement, ensuring improved traceability.

  • A direct link has been added to the Measurement Details page.

  • This link redirects to the Configuration Page of the device that captured the measurement, improving navigation efficiency.

v1-21_image4

GUI / Legacy Report – Updated Trivial Name

The trivial name of Phaseolus vulgaris has been updated:

  • Previous: White bean
  • New: Common bean Note: The allergen source material itself remains unchanged.

MAX 9k / MAX 45k – Monthly Cleaning Procedure

Improvement in Step 6 (Open Top Cover): During the monthly cleaning workflow, the Open Top Cover command now bypasses the initialization sequence.

  • The top cover opens immediately, reducing waiting time and streamlining the cleaning process.

On-Premise – QR Code Replacement Phased Out

    The standalone QR Code Replacement procedure has been outdated.
  • This functionality is now integrated into the Server Configuration Package update, which ensures all critical modules are updated consistently.
  • The synchronization package includes updates for:
    • QR Code Replacements
    • Contracts
    • Minimum Agent Required Version
    • Minimum MAX Firmware Required Version
    • QC Lots
    • Definitions
    • Run Configuration Templates
v1-21_image6

How to apply: Menu ➡️ System Admin ➡️Server Configuration ➡️Import Server Configuration

⚠️ Mandatory update: Due to this change, upgrading to RAPTOR SERVER v1.21 is required for all on-premise installations to ensure compatibility and proper synchronization.

Full List of Changes

RAPTOR SERVER GUI

  • QC evaluation logic unified across New Measurements and Device QC Measurements.
  • Stricter sample barcode validation (post-run barcode changes no longer classify as QC).
  • Reprocessed QC keep original capture timestamp for ordering in QC lists and charts.
  • QC result override possible (MADx only).
  • QC charts export: download individual charts as images and copy numerical datapoints.
  • Trend View CV fix: CV now calculated as (SD / Mean) × 100%.
  • Filters on new measurements/approved measurements page are now ignored in case a measurement is opened via a link e.g run overview page.
  • The 2025 report can now display the device information.
  • 2025 Report: device info (device type + serial number) can be printed.
  • 2025 Report (ALEX Food/Air): total IgE now shown as < 20 kU/L.
  • 2025 Report (ALEX Air): section title renamed from “Insects & Venoms” to “Insects”.
  • Report templates: templates can be duplicated.
  • 2025 Report formatting: HTML removed from address block and footer; use line breaks (Enter) only.
  • Trivial name update: Phaseolus vulgaris changed from White bean to Common bean in ALEX¹ and ALEX² (source material unchanged).
  • Automatic QC thresholds adjusted for ALEX¹, ALEX², and ALEX³.
  • Email sender address for notifications set to no-reply@madx.com
  • RAVEN1 patient symptoms are now translated.
  • Measurement source label has been improved.
  • Direct link to device configuration added on the Measurement Details page.
  • ASTM export now sets the correct RAPTOR SERVER version.
  • Country imprint (Serbia) updated.

MAX 9k / 45k

  • Monthly cleaning – Step 6 (Open Top Cover): skips initialization; cover opens immediately.
  • Run configuration templates updated.
  • Stop states now include more detailed descriptions.

On premise

  • PDF Service v2 default port changed from 8080 to 80.
  • Server Configuration Synchronization package introduced (replaces standalone QR Code Replacement).